Choosing an NDT Calibration Laboratory: Scope, Results and Traceability

A calibration sticker is a useful reminder, but it is not the complete evidence. When selecting a laboratory, start with the measurand, range and uncertainty your inspection work needs. Then check whether the proposed calibration service covers those requirements.

Read the accreditation scope

ISO/IEC 17025 addresses the competence and consistent operation of testing and calibration laboratories. An accreditation claim should be checked against the laboratory’s current published scope and the actual service ordered. Accreditation for one measurement does not establish competence for every instrument or range.

Specify the instrument identity, accessories, channels and functions that matter to your use. Agree whether adjustments are permitted and whether the report will distinguish the condition received from the condition returned.

Review the report before release

Check the equipment identifier, results, units, measurement conditions and uncertainty information relevant to the calibration. If a statement of conformity is requested, agree the specification and decision rule with the laboratory. An unexplained “pass” can leave the inspection team unable to judge suitability.

If the instrument was found outside the required tolerance, assess the effect on previous work through the quality system. Replacing the sticker does not answer that retrospective question.

Keep calibration and field checks separate

Laboratory calibration does not eliminate the equipment checks required by the NDT procedure. Retain those job checks with the instrument, probe or other relevant assembly used during the examination. Review recall intervals using applicable requirements, usage and performance history instead of assuming that one interval suits every item.

References and next steps

ISO/IEC 17025: laboratory competence requirements provides further background. Use the governing specification and approved procedure for the actual job.

Review calibration record management. Identify the instrument, measurement functions, ranges and calibration records your inspection procedure requires.

Where Atlantis can help

Atlantis provides NDT inspection, ASNT SNT-TC-1A training, Level III consulting and NDT software.